Partial hand prosthetic care requires more than selecting a terminal device. Successful outcomes depend on a repeatable method for evaluating residual anatomy, capturing an accurate impression, modifying the model, designing intrinsic suspension, and aligning digit terminal devices to support functional grasp.
This technical demonstration will be built upon prior partial hand socket design instruction by connecting interface principles to terminal device selection, mounting strategy, fabrication workflow, and prosthesis delivery. The session will review six design principles: palmar platform, thenar hook, hypothenar hook, metacarpal cap, ulnar window, and dorsal desert. These principles will be used to show how load is directed toward pressure-tolerant anatomy, how suspension can be achieved distal to the wrist, and how trimlines influence comfort, donning, and functional durability.
Live demonstration will include partial hand model review, modification planning, selected plaster modification techniques, diagnostic socket considerations, and terminal device alignment decision-making. Where appropriate, emerging lower-cost and resource-adapted approaches will be discussed as design considerations rather than product-specific claims.
The session is intended for clinicians and technicians seeking a practical framework for partial hand prosthetic fabrication across varied clinical environments. Attendees will leave with a structured method for translating evaluation findings into socket design features, alignment decisions, and fabrication plans that support comfort, function, durability, and access.
Statement of the objective / learning objectives
Attendees will learn a structured, clinically applicable method for evaluating partial hand presentations, designing sockets with intrinsic to the hand suspension, aligning mechanical digits, and adapting fabrication strategies to improve comfort, function, durability, and access.